Available Options
The 918D-ST-SL is a newly redesigned thin-profile, wand detector with a sturdy metal housing. Features include integrated calibration-data storage, built-in OD3 attenuation filter with automatic On/Off detection, and temperature sensing electronics for temperature drift compensation. Incorporating a large area 10 mm X 10 mm Silicon photodiode, this slim optical sensor is ideal for laser and other optical power measurements in tight locations with busy setups. It is supplied with a NIST traced calibration report that details individual detector responsivity measured with and without attenuator. Includes a DB15 calibration module that provides direct compatibility with Newport’s latest power meters, including 1916-R, 843-R, 1830-R, 1918-R, and 1936/2936-R. RoHS compliant.
feature | |
Attenuator | OD3, Built-in |
Calibration Uncertainty With Attenuator | 1% @ 400-940nm4% @ 941-1100nm |
Calibration Uncertainty without Attenuator | 1% @ 400-940nm4% @ 941-1100 nm |
Connector Type | DB15 |
Detector Type | Wand Style |
Linearity | ±0.5 % |
Material | Silicon |
Maximum Measurable Power | 2.0 W |
Maximum Measurable Power Without Attenuator | 4 mW with 1830-R and 1936-R, 2.5 mW with 843-R, 1919-R, and 841-PE-USB |
Maximum Power Density | 30 W/cm2 |
Maximum Power Density without Attenuator | 3 W/cm2 |
Maximum Pulse Energy with Attenuator | 5 µJ |
Maximum Pulse Energy Without Attenuator | 5 nJ |
Minimum Detectable Power | 20 pW with 1936-R/2936-R, 100 pW with 1830-R, 5 nW with 1918-R, 1 nW with 843-R & 1919-R |
Noise Equivalent Power | -70 dBm |
Operating Temperature Range | 5 to 50°C, <70% RH |
Rise Time | ≤3 µs |
RoHS | Compliant |
Sensor Size | 10 x 10 mm |
Spectral Range | 400 - 1100 nm |
Uniformity | ±2 % |