Available Options
The 819D-SL-3.3-CAL is a NIST traceable calibrated integrating sphere detector using a 3.3 inch integrating sphere with a high performance silicon photodetector. The input port size is 0.5 in. diameter and the wavelength range is 400 – 1100 nm. It comes with an SMA fiber optic connector on the North pole as a standard feature allowing a small amount of light pickoff for wavelength measurement or any further analysis without affecting the overall system calibration. Note that the system calibration is no longer valid if any component is changed from the original calibrated configuration. For a very high power level, elevated temperature of the integrating sphere system can affect the measurement accuracy so the sphere must be temperature controlled. Calibration is done only with no attenuator on the photodiode. If you have a legacy optical power meter that requires an 8-pin DIN calibration module, order 819D-SL-3.3-CAL/CM.
feature | |
Calibration Uncertainty without Attenuator | 1.5% @ 400-440 nm1% @ 441 – 950 nm1.5% 951 - 1000 nm3% 1001 - 1100 nm |
Connector Type | DB15 |
Detector Type | Sphere Detector |
Input Port Size | 0.5 in. |
Material | Silicon |
Power Range | 100 nW - 180 mW |
Sensor Size | Ø11.3 mm |
Spectral Range | 400 - 1100 nm |
Sphere Size | 3.3 in. |