Available Options
The LBP2-HR-IR2 Laser Beam Profiler, replacing the obsoleting camera of LBP2-HR-IR, is a powerful software driven camera with comprehensive beam diagnostic measurement features. It features the new Model SP928-1550, a 1/1.8 inch format 1924 x 1448 pixel CCD camera with phosphor coating for near IR sensitivity between 1440 and 1605 nm. The protective window on the CCD chip is removed to eliminate optical interference. Includes a USB3.0 compatible cable. The easy to use graphical user interface includes all of the accuracy and ISO approved quantitative results. Each laser beam profiler comes with one stackable LBP2-ND1 and two LBP2-ND2 filters with a 19 mm clear aperture, 1 inch-32UN (C-Mount). The damage threshold for these ND filters is 50 W/cm2, however, distortion due to thermal lensing can happen at as low as 5 W/cm2.
feature | |
Beam Size Accuracy | ±2 % |
CE | Compliant |
Computer Interface | USB3 |
Damage Threshold | 50 W/cm2 , 0.1 J/cm2 with all filters installed for < 100 ns pulse |
Detector Type | Si CCD with phosphor coating |
Dimensions | 48 x 44 x 20.2 mm |
Display Refresh Rate | 13 fps at full resolution (12 bit mode) |
Dynamic Range | 30 dB |
Gain | 0 dB to 24 dB |
Minimum Power Density | 50 µW/cm2 |
Minimum Spot Size | 600 µm |
Number of Pixels | 1928 x 1448 |
Operating Mode | CW and Pulsed, Interline transfer progressive scan CCD |
Pixel Spacing | 3.69 x 3.69 µm |
RoHS | Compliant |
Saturation Intensity | 7 mW/cm2 @ 1550 nm |
Sensor Size | 7.1 x 5.3 mm |
Shutter | 30 µs to multiple frame times |
Spectral Range | 1440-1605 nm |