National Instruments TC-2095 Rack-mount thermocouple terminal block
1 image: Click to enlarge
National Instruments · SCXI

TC-2095

The NI TC-2095 connects 32 thermocouples to compatible SCXI modules through a rack-mount terminal block with cold-junction sensing.

ManufacturerNational Instruments
ModelTC-2095
Part Number777509-01
Product TypeRack-mount thermocouple terminal block
● Available to Order
Condition: confirmed per unit

Request current pricing and lead time for this model. Our team will confirm the exact equipment configuration and condition.

Fast RFQ

Quote: TC-2095

We will reply by email.
Free Shippingfull policy linked
30-Day Warrantystandard coverage
30-Day Returnspolicy applies
Technical Supportmodel & compatibility help
Connections32 uncompensated thermocouple miniconnectors
Mounting19-inch rack
Module connection96-pin DIN
SOHO Assistant
Ask about this exact model, a replacement, accessory, driver, rental or repair question.
Ask about specifications, compatibility, availability, rental or repair.
Ask SOHO Assistant
Overview

Product overview

Key product information and technical features.

The TC-2095 provides thermocouple connections for SCXI-1100 and SCXI-1102/B/C modules. Rear switches configure signal referencing independently for each channel.

  • Isothermal construction reduces temperature gradients across thermocouple junctions.
  • Integrated thermistor senses cold-junction temperature.
  • Switchable pull-up resistors support open-thermocouple detection.
Verified specifications

Technical data

Verified product specifications.

Thermocouple connections32 channels
Connector contactsCopper; uncompensated
Cold-junction sensor output1.91 at 0 °C; 0.58 at 55 °C V
Cold-junction sensor accuracy with SCXI-1102/B/C0.5 over 15–35 °C °C
Cold-junction sensor accuracy with SCXI-11000.65 over 15–35 °C °C
Condition & included items

What the actual unit includes

Confirm condition, included accessories and revision before ordering.

Condition

Confirmed per unit.

Included

Device / accessories confirmed in quote.

Revision

Exact revision confirmed when relevant.

Compatibility

Complete the measurement chain

Review system components and compatibility requirements.

National Instruments

Manufacturer

SCXI

Signal-conditioning platform

TC-2095

Thermocouple connection terminal block

Connection cables

SH96-96 shielded cable or R9696 ribbon cable with TBX cable adapter.

Check grounding and resistor-network settings before wiring; protect the terminal block from extreme temperature differentials.
Manuals & software

Technical resources

Manuals, software and technical resources for this model.

Shipping, warranty & returns

Commercial terms

Review shipping, warranty and return terms before ordering.

Shipping
Free Shipping. Full policy terms apply.
Warranty
30-Day Warranty. Standard coverage applies.
Returns
30-Day Returns. Return policy applies.
Price / Availability
Contact SOHO ProLab for current quote and availability.
Lifecycle services

Source, rent or repair the exact model

Choose pricing, rental or repair for this exact model.

FAQ

Product-specific questions

Product-specific answers based on available technical information.

Which SCXI modules are recommended?

NI lists SCXI-1100 and SCXI-1102/B/C. Its terminal-block datasheet does not recommend TC-2095 use with SCXI-1104/C.

Are there thermocouple-type precautions?

Yes. NI warns that copper connector contacts can introduce offset errors with T-, R-, and S-type thermocouples.