NI RM-26999 is a high power rackmount terminal block that NI also describes as a rack-mounted signal conditioning device for power measurements. Official NI sources identify RM-26999 as a primary hardware module rather than a cable or accessory, and the reviewed NI product page shows a single orderable part number, 786328-01. NI positions the unit for connection to simultaneous-sampling multifunction I/O devices for power measurement setups.
The RM-26999 provides 4 voltage input channels and 4 current transducer ports. NI states that the unit does not directly measure current; instead, it works with current transducers and supports transducer connections that include ±10 V signal lines, +15 V and -15 V transducer supply outputs, and calibration/status lines. On the DAQ side, NI specifies supported current input ranges of ±1 V, ±2 V, ±5 V, and ±10 V.
For measurement performance, NI specifies 200:1 attenuation, typical DC common-mode rejection greater than 100 dB, single-ended-to-earth input impedance of 10 MΩ || 4.7 pF, and 1 MHz bandwidth at -3 dB. Additional published specifications include burden resistor selections of 0.5 Ω, 1 Ω, 2 Ω, 5 Ω, and 10 Ω, plus shunt accuracy of ±0.05% using a metal foil shunt rated to 2 W maximum. Maximum voltage ratings are 1,000 V Category II and 2,000 V peak for other non-MAINs circuits.